Description
Channel & Wiring Configuration
-
8-channel fully differential independent analog input
-
Per-channel standalone filtering and amplification circuit (no shared circuitry)
-
All channels multiplexed to one DAQ analog input channel
-
Front 96-pin DIN connector, rear 50-pin SCXI backplane interface
Programmable Gain & Input Range
-
Software-selectable gain multiples: ×1, ×2, ×5, ×10, ×20, ×100
-
Maximum raw input range: ±5 V differential voltage
-
Minimum measurable input: ±50 mV at ×100 gain
-
Default input coupling: DC coupling (AC coupling requires SCXI-1304 accessory)
On-board Filter Parameters
-
Filter type: 8th-order elliptic low-pass anti-aliasing filter
-
Fixed cutoff frequency: 12.5 kHz
-
Stopband attenuation: 120 dB for frequencies above 13.5 kHz
-
Max channel scan rate: 50 kS/s for multi-channel synchronous scanning
Key Electrical Specifications
-
Input impedance: 10 GΩ parallel with 40 pF
-
60Hz CMRR: >108 dB
-
Typical measurement accuracy: ±0.01% of reading
-
Overvoltage protection: ±30 VDC powered on, ±15 VDC powered off
-
Channel crosstalk: <-92 dB at 1 kHz
-
Power consumption: 3.1 W
Mechanical & Environmental Specifications
-
Dimension: 172×86×32 mm, standard SCXI single-slot size
-
Operating temperature: 0℃ ~ +55℃
-
Storage temperature: -40℃ ~ +85℃
-
Humidity: 10%–90% non-condensing
Software Compatibility
-
Drivers: NI-DAQmx, Traditional NI-DAQ
-
Supported software: LabVIEW, LabWindows/CVI, Python DAQmx, C#
-
Perfect matched terminal block: SCXI-1304 AC/DC coupling terminal block
Pros & Limitations
-
Pros: Per-channel independent anti-aliasing filter, ultra-high input impedance, excellent high-frequency noise suppression, low measurement drift for dynamic signals
-
Limitations: Fixed filter cutoff frequency, no built-in channel isolation, no internal bridge excitation for strain sensors








Reviews
There are no reviews yet.