Description
Channel & Relay Configuration
-
16 independent SPDT latching relays (Form C), no shared channel circuitry
-
Each channel contains Common, Normally Open (NO), Normally Closed (NC) terminals
-
Rear 96-pin shielded DIN connector for front terminal block connection
-
All relays controlled via SCXI backplane digital trigger bus
Relay Electrical Ratings
-
Max switching voltage: 150 VDC, 125 VAC RMS
-
Max switching current: 2 A per channel; continuous carry current: 5 A
-
Max switching power: 60 W (DC), 62.5 VA (AC)
-
Contact on-resistance: ≤50 mΩ initial value
Switching & Lifespan Parameters
-
Maximum switching rate: 115 channel operations per second
-
Mechanical relay lifespan: 100 million cycles (no-load switching)
-
Electrical relay lifespan: 200,000 cycles at rated load
-
Operating actuation time: 2.5 ms typical, 5 ms maximum
Key Electrical Specifications
-
Channel-to-channel isolation: 250 Vrms continuous isolation
-
Transient surge withstand: 1000 V 1μs lightning surge
-
Thermal offset: <1 μV for low-level analog signal switching
-
Backplane power consumption: 2.9 W typical operating power
-
Contact material: Gold-plated ruthenium alloy for anti-oxidation
Mechanical & Environmental Specifications
-
Dimension: 172×86×32 mm, standard single-slot SCXI module
-
Operating temperature: -20℃ ~ +70℃
-
Storage temperature: -40℃ ~ +85℃
-
Humidity: 10%–90% non-condensing
Compatibility Matching List
-
Matched terminal block: SCXI-1324 high-voltage screw terminal block
-
Compatible chassis: SCXI-1000, SCXI-1001
-
Compatible controllers: SCXI-1600, PCI/PXI SCXI adapters
-
Official status: Obsolete, available for aftermarket replacement
Software Compatibility
-
Drivers: NI-SWITCH, NI-DAQmx, Traditional NI-DAQ
-
Configuration tools: NI MAX, NI Switch Executive
-
Supported software: LabVIEW, LabWindows/CVI, Python, C#
Pros & Limitations
-
Pros: Latching relays with zero holding power consumption, ultra-low thermal offset for weak signal switching, high channel isolation, long mechanical service life, mature matching with SCXI-1324
-
Limitations: Obsolete official product, no built-in surge protection, lower switching speed than solid-state relays, periodic contact maintenance required