Description
Channel & Topology Configuration
-
32 single-ended RF signal channels with four dedicated common (COM) ports
-
Predefined standard topologies: 8×1, 16×1, 32×1 unterminated multiplexers
-
Custom flexible topologies: 4×1/8×1 externally terminated multiplexers, 36-terminal sparse matrix
-
Quad independent 8-channel signal banks for parallel multi-group RF switching
-
Rear 96-pin shielded DIN connector for SCXI chassis backplane communication
Core RF Electrical Specifications
-
Characteristic impedance: Fixed 50 Ω for standard RF coaxial matching
-
Operating bandwidth: DC to 500 MHz (-3 dB insertion loss threshold)
-
Insertion loss: ≤0.3 dB @ 100 MHz, ≤0.8 dB @ 500 MHz per signal path
-
VSWR: ≤1.2:1 across full bandwidth for minimal signal reflection
-
Channel isolation: >70 dB @ 100 MHz, >52 dB @ 500 MHz
Relay Switching Ratings & Timing
-
Switching signal power: Max 1 W continuous RF power per channel
-
DC current rating: 0.5 A switching current, 1 A continuous carry current
-
Relay type: Break-before-make electromechanical RF relays
-
Operate/release time: 4 ms typical total switching latency
-
Max scan rate: 100 channel switching operations per second
-
Phase matching: <100 ps inter-bank propagation skew
Reliability & Isolation Parameters
-
Mechanical relay lifespan: 100 million no-load switching cycles
-
Electrical lifespan: 2 million cycles under 0.5 W RF load
-
Channel-to-chassis isolation: 1000 Vrms transient surge withstand
-
Initial on-path resistance: <1.0 Ω, end-of-life threshold ≥2.0 Ω
-
Internal stub suppression: Optimized dual-path routing to eliminate RF reflection
Mechanical & Environmental Specifications
-
Dimension: 172×86×32 mm, standard single-slot SCXI module footprint
-
Operating temperature: 0℃ ~ +50℃
-
Storage temperature: -20℃ ~ +70℃
-
Humidity: 10%–90% non-condensing, max working altitude 2000 m
-
EMC compliance: CE, FCC Class A industrial electromagnetic certification
Compatibility Matching List
-
Matched terminal block: SCXI-1334 50 Ω RF coaxial terminal block
-
Compatible chassis: SCXI-1000, SCXI-1001 AC-powered chassis
-
Compatible controllers: SCXI-1600, PCI/PXI SCXI bus adapters
-
Official status: Obsolete, no new official inventory supply
Software Compatibility
-
Drivers: NI-SWITCH 2.0 and above, NI-DAQmx, Traditional NI-DAQ
-
Configuration tools: NI MAX, NI Switch Executive
-
Supported software: LabVIEW, LabWindows/CVI, Python DAQmx, C#
Pros & Limitations
-
Pros: 500 MHz wide RF bandwidth, flexible software-defined topologies, low phase skew for differential signals, low insertion loss, standardized 50 Ω impedance matching, long relay service life
-
Limitations: No onboard RF terminators (requires external accessories), limited RF power tolerance, obsolete lifecycle, not compatible with low-frequency analog signal multiplexing







Reviews
There are no reviews yet.