Description
Channel Configuration
-
8-channel fully differential isolated analog input
-
Independent channel-to-channel & channel-to-ground galvanic isolation
-
All channels multiplexed to one DAQ analog input channel
Programmable Input Parameters
-
Input voltage range: ±10 V max raw input
-
Programmable gain multiples: ×1, ×2, ×5, ×10, ×20, ×50, ×100, ×200, ×500, ×1000 (12 preset gears)
-
Min measurable input: ±10 mV (at ×1000 gain)
On-board Low-pass Filter Parameters
-
10 kHz 4th-order low-pass filter: Maximum scan rate 10 kS/s
-
4 Hz 4th-order low-pass filter: Maximum scan rate 3 S/s for 50/60Hz industrial power frequency noise rejection
-
No full-bandwidth bypass mode (fixed two-filter structure)
Key Electrical Specifications
-
Input impedance: 1 GΩ ultra-high impedance
-
60Hz CMRR: >115 dB
-
Nonlinearity: 0.002% FSR
-
Continuous isolation withstand: 300 Vrms channel-to-channel/channel-to-ground
-
Transient overvoltage protection: 1000 V surge
-
Power consumption: 2.8 W
-
Built-in CJC sensor for multi-type thermocouple measurement
Mechanical & Environmental Specifications
-
Front connector: 96-pin DIN shielded connector
-
Dimension: 172×86×32 mm
-
Operating temperature: -40℃ ~ +85℃
Software Compatibility
-
Drivers: NI-DAQmx, Traditional NI-DAQ
-
Supported software: LabVIEW, LabWindows/CVI, Python, C#
Pros & Limitations
-
Pros: 300 Vrms high isolation, ultra-high input impedance, built-in thermocouple cold junction compensation, compact channel-independent filtering
-
Limitations: No frequency input function, only two filter options, low synchronous scanning performance for dynamic signals








Reviews
There are no reviews yet.